FDA recall Z-0927-2018

Medtronic Navigation, Inc.-Littleton · Class II · device

Product

O-arm¿ O2 interventional fluoroscopic x-ray system; Catalog Number: B1-700-02000 The O-arm O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and pediatric patients weighting 60 lbs or greater and having an abdominal thickness of greater than 16cm, and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The O-arm O2 Imaging System is compatible with certain image guided surgery systems.

Reason for recall

During servicing of certain internal components of the O-Arm O2 Surgical Imaging System, it is possible that the high-power cable connector may come in contact with surrounding metal during the disconnecting/reconnecting process which may result in an electrical short and a shock to the field service technician.

Distribution

Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico and to the states of : AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI, WV., and to the countries of : Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Finland, France, Germany, Iceland, India, Ireland, Israel, Italy, Japan, Jordan, Korea, Luxembourg, Martinique, Mexico, Netherlands, New Zealand, Poland, Portugal, Qatar, Romania, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, and United Kingdom.

Key facts

Status
Terminated
Initiation date
2017-09-30
Report date
2018-03-14
Termination date
2019-05-02
Voluntary/Mandated
Voluntary: Firm initiated
Location
Littleton, MA, United States