FDA recall Z-0939-2022

Graphic Controls Acquisition Corporation · Class II · device

Product

CLARAVUE Pre-wired Electrode - IP SET 6V RT ADULT-US Model: 32028780 50603-US

Reason for recall

Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard requires the device to withstand, without breakdown , an application of, 5,000 V d.c. when used with a defibrillator at a voltage above its maximum threshold, Claravue does not meet the specification and delivers an inefficient electrical charge which may result in an ineffective defibrillation.

Distribution

US Nationwide distribution in the States of AR, CA, CO, FL, IN, KS, MA, MN, NY, OK, VT.

Key facts

Status
Terminated
Initiation date
2022-03-15
Report date
2022-04-20
Termination date
2024-07-12
Voluntary/Mandated
Voluntary: Firm initiated
Location
Buffalo, NY, United States

Primary source