FDA recall Z-0944-2019

Arrow International Inc · Class II · device

Product

0F 550 x 8ML/HR 2ML 60LO CPNB, Product code MT8060XL CPNB, Arrow AutoFuser Disposable Pain Control Pump

Reason for recall

Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.

Distribution

Nationwide distribution.

Key facts

Status
Terminated
Initiation date
2018-10-02
Report date
2019-03-06
Termination date
2021-11-12
Voluntary/Mandated
Voluntary: Firm initiated
Location
Reading, PA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0944-2019