FDA recall Z-0946-2021

Heartware, Inc. · Class I · device

Product

Medtronic HVAD Pump Implant Kits, labeled as: a. PUMP 1103, b. PUMP 1104, c. PUMP 1104JP, d. PUMP 1100, e. Model 1101, f. Model 1102, g. Model 1104CA-CLIN, h. Model 1205, i. Model MCS1705PU

Reason for recall

Pumps manufactured with impellers from a subset of lots from a single supplier are failing to initially start, restart, or experiencing a delay in restart within the maximum number of restart attempts at a rate substantially higher than pump in the overall population.

Distribution

Worldwide - US Nationwide

Key facts

Status
Ongoing
Initiation date
2020-11-19
Report date
2021-02-10
Voluntary/Mandated
Voluntary: Firm initiated
Location
Miami Lakes, FL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0946-2021