FDA recall Z-0962-2022

Siemens Medical Solutions USA, Inc · Class II · device

Product

SIEMENS SOMATOM with software syngo CT VA30A_SP4 Models: SOMATOM go.Now 11061610 & 11061618 SOMATOM go.Up 11061620 & 11061628 SOMATOM go.All 11061630 & 11061638 SOMATOM go.Top 11061640 & 11061648 SOMATOM go.Sim 11061660 & 11061668 SOMATOM go.Open Pro 11061670 & 11061678 SOMATOM X.cite 11330001 Intended Use: Computed tomography x-ray system

Reason for recall

Software error may result in sporadic problems causing scanning workflow interruptions, unexpected user notifications and image artifacts. Sporadic software errors may also occur during interventional workflows. Results in potential patient issues: Possible rescan , Unexpected X-Ray dose and additional contrast media, Delay in diagnosis, scan abort,and patient rescan

Distribution

Nationwide

Key facts

Status
Ongoing
Initiation date
2022-03-07
Report date
2022-04-27
Voluntary/Mandated
Voluntary: Firm initiated
Location
Malvern, PA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0962-2022