# FDA recall Z-0964-2026

> **Intuitive Surgical, Inc.** · Class II · device recall initiated 2024-08-21.

## Product

da Vinci 5, Surgeon Console Column Motor Connector; ASSY,DV5 CONSOLE,SSC,IS5000; Part Number: 380730-40

## Reason for recall

Surgical system Error 23062 due to connector1) Partially damaged "Potential problem with ergonomics: Disable - Retry " or 2) Failed "Disable Ergonomics: Viewer, headrest, armrest, and pedal tray position cannot be adjusted until da Vinci is restarted. Contact Technical Support to report the issue: Cancel-Disable" ergonomics locked until motor replaced; may lead to delay/aborted/alternate procedure

## Distribution

US Nationwide distribution in the states of MN, FL, CO, CA, NC, KS, TX, SC, WV, LA, OH, NY, NH, GA, KY, OK, NJ, MA, OR, PA, NM, AZ, TN, ND, IL, AR, WA, IN, VA, CT.

## Key facts

- **Recall number:** Z-0964-2026
- **Recalling firm:** Intuitive Surgical, Inc.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2024-08-21
- **Report date:** 2025-12-31

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Sunnyvale, CA, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0964-2026

## Citation

> AI Analytics. FDA recall Z-0964-2026. Retrieved 2026-07-26 from https://api.ai-analytics.org/recall/Z-0964-2026. Source: US FDA. Licensed CC0.

---