FDA recall Z-0968-2018

Becton Dickinson & Company · Class II · device

Product

BD Vacutainer¿ K2 EDTA (K2E) 3.6mg Blood Collection Tube (13 x 75 mm x 2.0 mL). Catalog Number: 367841. Used for whole blood hematology determinations, immunohematology testing and blood donor screening.

Reason for recall

A limited portion of the lot was manufactured with less than the required amount of K2EDTA additive, an anticoagulant, which may cause erroneous results that could lead to specimen recollection and potential delay of treatment or misdiagnosis/mismanagement of treatment and serious complications, such as failure to detect thrombocytosis or postponing surgery.

Distribution

Nationwide Distribution

Key facts

Status
Terminated
Initiation date
2017-08-29
Report date
2018-03-21
Termination date
2019-05-16
Voluntary/Mandated
Voluntary: Firm initiated
Location
Franklin Lakes, NJ, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0968-2018