FDA recall Z-0971-2019

Abbott Laboratories, Inc · Class II · device

Product

ARCHITECT MAGNESIUM, (a) REF 7D70-21, (b) 7D70-31 The Magnesium assay is used for the quantitation of magnesium in human serum, plasma, or urine.

Reason for recall

Abbott internal testing has identified that the Magnesium urine application demonstrates depressed urine result recovery. All samples >0.5mEq/L fail the linearity acceptance criteria for the bias specification and demonstrate depressed recovery of up to 37%. The specific cause of the negative bias/depressed results is currently under investigation.

Distribution

Worldwide Distribution: US (nationwide) and to countries of: CHINA, THAILAND, GUATEMALA, and EL SALVADOR.

Key facts

Status
Terminated
Initiation date
2019-01-02
Report date
2019-03-13
Termination date
2021-06-30
Voluntary/Mandated
Voluntary: Firm initiated
Location
Irving, TX, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0971-2019