FDA recall Z-0973-2019

Stryker Sustainability Solutions · Class II · device

Product

Stryker Sustainability Solutions (SSS) reprocessed Arthroscopic Shavers Arthroscopic shavers can be used to abrade, cut and excise tissue and bone; remove loose fragments; and, shave away debris in arthroscopic surgeries, as well as surgeries of the jaw and sinuses. Intended for use in orthopedic surgical procedures of the joints, jaw or sinuses where the cutting and removal of soft and hard tissue or bone is needed in patients requiring orthopedic surgery.

Reason for recall

Affected lots of reprocessed arthroscopic shavers may display the incorrect catalog number and speed settings when connected to the generator, require manual mode for use on CORE generators, or display an error message on Crossfire 1 and 2 generators which does not allow the device to be used.

Distribution

Nationwide Distribution: AL, AR, AZ, CA, CO, DC, DE, FL, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY and Puerto Rico. OUS Israel and Canada.

Key facts

Status
Terminated
Initiation date
2018-07-27
Report date
2019-03-13
Termination date
2020-09-17
Voluntary/Mandated
Voluntary: Firm initiated
Location
Tempe, AZ, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0973-2019