FDA recall Z-0975-2024

Philips North America · Class II · device

Product

Ingenia Ambition X (Product Number 781356) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system

Reason for recall

Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.

Distribution

Worldwide - US Nationwide distribution in the states of FL and NY. The countries of Denmark, Norway, and Sweden.

Key facts

Status
Ongoing
Initiation date
2022-01-06
Report date
2024-02-14
Voluntary/Mandated
Voluntary: Firm initiated
Location
Cambridge, MA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0975-2024