FDA recall Z-0979-2018

Zimmer Biomet, Inc. · Class II · device

Product

Humeral Stem 83mm(Size 17) Product Usage: Intended for primary and revision total shoulder arthroplasty as well as hemi-arthroplasty

Reason for recall

Zimmer Biomet is conducting a medical device recall for two lots of Comprehensive Humeral stems. The lots were commingled during the manufacturing process and were etched with incorrect lot numbers. The size difference between the two stem sizes is 1.1 inches. It is visually recognizable by the user that the box label does not match the contents of the box.

Distribution

US Nationwide in the states of MN, NJ, CA, TX, ND

Key facts

Status
Terminated
Initiation date
2017-12-08
Report date
2018-03-21
Termination date
2018-09-04
Voluntary/Mandated
Voluntary: Firm initiated
Location
Warsaw, IN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0979-2018