# FDA recall Z-0981-2024

> **Beckman Coulter, Inc.** · Class II · device recall initiated 2023-11-21.

## Product

Access Ultrasensitive Insulin only used on the DxI 9000 Access Immunoassay Analyzer, Catalog Number 33410

## Reason for recall

There is an issue with the Access Ultrasensitive Insulin assay protocol file (APF) for use with the DxI 9000 Access Immunoassay Analyzer (APF 182 test definition version 6.7 and below).  The Access Ultrasensitive Insulin APF applies an incorrect conversion factor when converting from system default units of ¿IU/mL to the International System of Units (SI units) of pmol/L.

## Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Croatia, Czech Republic, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, United Kingdom.

## Key facts

- **Recall number:** Z-0981-2024
- **Recalling firm:** Beckman Coulter, Inc.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2023-11-21
- **Report date:** 2024-02-14

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Chaska, MN, United States