# FDA recall Z-0987-2020

> **Care Essentials Pty., Ltd.** · Class II · device recall initiated 2020-01-07.

## Product

Cocoon Convective Warming System, Product Code: CWS5000 (110V).

## Reason for recall

Potential for thermal damage near the power cord connector and Power Entry Module (PEM) due to improper connection of the power cord connector and PEM pins on the Cocoon Convective Warming System.

## Distribution

US Nationwide in the states of  AZ CA FL GA KS LA MA MD MI MN NC NM NY OH PA QC SC TX VA WV and the Foreign distribution to Canada.

## Key facts

- **Recall number:** Z-0987-2020
- **Recalling firm:** Care Essentials Pty., Ltd.
- **Classification:** Class II
- **Product type:** device
- **Status:** Terminated
- **Initiation date:** 2020-01-07
- **Report date:** 2020-02-12
- **Termination date:** 2023-03-22

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** North Geelong, N/A, Australia

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0987-2020

## Citation

> AI Analytics. FDA recall Z-0987-2020. Retrieved 2026-06-29 from https://api.ai-analytics.org/recall/Z-0987-2020. Source: US FDA. Licensed CC0.

---