FDA recall Z-0987-2022

Siemens Medical Solutions USA, Inc · Class II · device

Product

Artis pheno- in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). Intended use-single and biplane diagnostic imaging and interventional procedures. Model: 10849000

Reason for recall

Five potential software issues affecting Artis pheno and Artis icono systems in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). The potential software errors are related to the block movements function and manual rotation of the flat detector. Depending on the status of the intervention, the limited functionality may not be sufficient to continue with treatment as planned. This may result in a situation where clinical treatment may be delayed until an additional restart is performed

Distribution

US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2022-02-25
Report date
2022-04-27
Voluntary/Mandated
Voluntary: Firm initiated
Location
Malvern, PA, United States