FDA recall Z-0992-2019

Ortho-Clinical Diagnostics · Class II · device

Product

VITROS Chemistry Products Na+ Slides, Catalog Number 8379034 For in vitro diagnostic use only. VITROS Na+ Slides quantitatively measure sodium (Na+) concentration in serum, plasma, and urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

Reason for recall

Imprecision in the affected lot may produce negatively or positively biased patient sample and quality control fluid results.

Distribution

US Distribution to states of: AL, CA, CT, FL, HI, IA, ID, IL, LA, MI, MO, MS, NC, NM, NY, OK, PA, TX, WV, and WY; and internationally to: Canada.

Key facts

Status
Terminated
Initiation date
2019-01-22
Report date
2019-03-20
Termination date
2022-06-10
Voluntary/Mandated
Voluntary: Firm initiated
Location
Rochester, NY, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0992-2019