FDA recall Z-1010-2018

Fresenius Medical Care Renal Therapies Group, LLC · Class II · device

Product

NaturaLyte¿ Liquid Acid Concentrate 2.0 mEq/L, Model # 08-2251-0 3.43 liter bottle This concentrate is formulated to be used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.

Reason for recall

The label specifies the potassium concentration is 2.0 mEq/L. Some of the product in the affected lot was found to have a potassium concentration of 2.25 mEq/L.

Distribution

US Distribution to states of: CA, FL, KS, LA, and TX.

Key facts

Status
Terminated
Initiation date
2017-09-29
Report date
2018-03-21
Termination date
2018-08-09
Voluntary/Mandated
Voluntary: Firm initiated
Location
Waltham, MA, United States