FDA recall Z-1014-2022

Instrumentation Laboratory · Class II · device

Product

ACL TOP 300 CTS/350 CTS/500 CTS/550 CTS/970 CL: Name Part Number 1.ACL TOP 300 CTS 00000280060 2. ACL TOP 300 CTS 00000280060R (Refurbished) 3. ACL TOP 350 CTS 00000280065 4. ACL TOP 500 CTS 00000280040 5. ACL TOP 500 CTS 00000280040R ( Refurbished) 6/ 6. ACL TOP 550 CTS 00000280045 7. ACL TOP 970 CL 00000280097

Reason for recall

Updated and revised user instructions issued 7/14/22. Potential carryover issue may cause elevated quality control and sample results for heparin, apixaban or rivaroxaban when HemosIL Liquid Anti-Xa reagent is used in the same run with both HemosIL Liquid Antithrombin and HemosIL Q.F.A. Thrombin (Bovine), specific to instrument models that utilize a single reagent probe

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belarus, Belgium, Bermuda, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Ecuador, El Salvador, France, Germany, Gibraltar, Great Britain, Greece, Guatemala, Guyana, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Korea, Kosovo, Kuwait, Lebanon, Lithuania, Luxembourg, Macau, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russia, S. Tome, Saudi Arabia, Serbia, Singapore, Slovakia, Spain, Switzerland, Taiwan, Tanzania, Thailand, Trinidad, Tobago, Tunisia, Turkey, United Arab Emirates, Uruguay, Venezuela, Vietnam.

Key facts

Status
Ongoing
Initiation date
2022-02-22
Report date
2022-05-04
Voluntary/Mandated
Voluntary: Firm initiated
Location
Bedford, MA, United States