FDA recall Z-1018-2023

Biomet, Inc. · Class II · device

Product

Oxford Partial Knee System, Fixed Lateral Tibial Construct- Intended to replace part of a knee joint when used in conjunction with a femoral and tibial component Item Number: 154341

Reason for recall

Packaged in the incorrect shelf carton. Incorrect shelf carton; outer carton is smaller than the required outer carton, which forces the Tyvex pouch to be folded and packaged in the carton causing compound creases and compromise the sterile barrier

Distribution

Worldwide distribution - US Nationwide distribution in the states of IN, KY and the countries of Canada, New Zealand, Japan, Netherlands.

Key facts

Status
Ongoing
Initiation date
2022-12-15
Report date
2023-02-01
Voluntary/Mandated
Voluntary: Firm initiated
Location
Warsaw, IN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1018-2023