# FDA recall Z-1024-2018

> **Zimmer Biomet, Inc.** · Class II · device recall initiated 2017-11-29.

## Product

VERSYS 6 INCH BEADED FC 14X160MM STD BODY EXT & STD NECK    Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented

## Reason for recall

The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.

## Distribution

Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Brazil, Bolivia, Canada, Chile, China, Colombia, Ecuador, El Salvador, India, Japan, Malaysia, Mexico, Netherlands, New Zealand, Nicaragua, Panama, Seoul-Korea, Singapore, Taiwan, Thailand, and UAE.

## Key facts

- **Recall number:** Z-1024-2018
- **Recalling firm:** Zimmer Biomet, Inc.
- **Classification:** Class II
- **Product type:** device
- **Status:** Terminated
- **Initiation date:** 2017-11-29
- **Report date:** 2018-03-21
- **Termination date:** 2020-04-02

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Warsaw, IN, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1024-2018