FDA recall Z-1040-2023

Siemens Medical Solutions USA, Inc · Class II · device

Product

System Material # Artis zee floor 10094135 Artis zee ceiling 10094137 Artis zee multi-purpose 10094139 Artis zee biplane 10094141 Artis zeego 10280959 Artis Q floor 10848280 Artis Q ceiling 10848281 Artis Q biplane 10848282 Artis Q.zeego 10848283 Artis Q.zen floor 10848353 Artis Q.zen ceiling 10848354 Artis Q.zen biplane 10848355 Fluoroscopic X-Ray System

Reason for recall

In rare cases, the system may only boot into backup mode after an abrupt shutdown and not reach full operating mode. This may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system.

Distribution

Worldwide - US Nationwide distribution.

Key facts

Status
Ongoing
Initiation date
2022-12-16
Report date
2023-02-08
Voluntary/Mandated
Voluntary: Firm initiated
Location
Malvern, PA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1040-2023