FDA recall Z-1041-2022

Philips North America Llc · Class II · device

Product

Philips Hemodynamic Application, Model Number 722463, Software Version Number: R1.2.2 and R1.2.3. For visualization and analysis of patient hemodynamic data in the Philips Interventional Hemodynamic System.

Reason for recall

When a Neonatal patient is admitted, the Non-Invasive Blood Pressure is not set to Neonatal mode on the Philips Xperflex Cardio FC2010 patient monitor. If the NIBP is not set to Neonatal, the cuff inflation pressure applied will be higher than 150 mmHg and the deflation pressure will be higher than 5 mmHg. An applied cuff inflation pressure higher than 150mmHg or a deflation pressure higher than 5 could lead to bruising, mmHg nerve damage or bone deformity for a neonate.

Distribution

Devices were distributed to the following US states: FL, ID, NJ, OK, SC and TX. Devices were distributed to the following foreign countries: France, Germany, India, Mexico, Netherlands, New Zealand, Norway, Poland, South Africa, and Switzerland.

Key facts

Status
Terminated
Initiation date
2022-02-22
Report date
2022-05-18
Termination date
2024-11-06
Voluntary/Mandated
Voluntary: Firm initiated
Location
Cambridge, MA, United States

Primary source