FDA recall Z-1053-2023

GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING · Class I · device

Product

GE Nuclear Medicine 600/800 series systems, specifically: a) 870, b) 870 DR 3/8 inch NM GANTRY WITH NEW ROTOR - MEA, c) 870 DR / Dismount, d) 870 DR / Dismount / 60HZ, e) 870 DR 3/8 inch WSO, f) 870 DR 3/8 inch NM Gantry, g) 870 DR 3/8 inch, h) 870 DR 3/8 inch CH, i) 870 DR 5/8 inch NM GANTRY, j) 870 DR BJ / Dismount, k) 870 DR CyberDefense 3/8 inch NM Gantry, l) 870 DR NM GANTRY FINAL - NEW ROTOR, m) 870 DR China, n) D670 NM GANTRY WITH NEW ROTOR, o) D670 DR 3/8 inch 16 SL, p) D670 DR 3/8 inch MEA, q) D670 DR 3/8 inch, r) D670 DR 3/8 inch 16SL, s) D670 DR 3/8 inch 16SL NR, t) D670 DR 3/8 inch BJ, u) D670 DR 3/8 inch MEA, v) D670 DR NM Gantry 3/8 inch, w) D670 DR NM Gantry 5/8 inch, x) D670 DR with new rotor, y) D670 ES NM Gantry 3/8 inch, z) D670 ES NM/CT 3/8 inch O540+ NM Gantry, aa) D670 ES NM/CT DR 3/8 inch, bb) D670 GANTRY, cc) D670 PRO NM Gantry 3/8 inch, dd) D670 PRO NM GANTRY 3/8 inch, ee) DISC NM/CT670 DR 3/8 inch, ff) DISC NM/CT870 DR 3/8

Reason for recall

A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.

Distribution

Worldwide distribution. US nationwide including Puerto Rico, Algeria, Argentina, Australia, Austria, Bahrain, Belgium, Bolivia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, El Salvador, Ethiopia, Finland, France, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kuwait, Libya, Lithuania, Malaysia, Martinique, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, Uzbekistan, and Vietnam.

Key facts

Status
Ongoing
Initiation date
2022-12-19
Report date
2023-02-15
Voluntary/Mandated
Voluntary: Firm initiated
Location
Tirat Carmel, Israel

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1053-2023