FDA recall Z-1055-2019

Philips North America, LLC · Class II · device

Product

MultiDiagnost Eleva w/ Flat Detector, Product codes 708034, 708037, 708038 Product Usage: As a multifunctional I universal imaging application system, General RIF, Fluoroscopy, Radiography and Angiography can be performed along with more specialized interventional applications on human patients. This includes the following general areas: Digestive system: Swallowing studies, Oesophagus, Stomach, Small intestine, Colon, Defeacography, ERCP, T-tube cholangiogram, Liver biopsies, Transjugular Intrahepatic Portosystemic Shunts (TIPS) Skeletal system: Bone studies. Urinary system: IVP, Cystograms, Percutaneous, Nephrolithotomy, Nephrostomy tube replacement Reproductive system: Hysterosalpingogram, Vena spermatica, Cavernography Respiratory system: Thorax, Bronchoscopy, Pulmonary biopsies Circulatory system: Venography, Arteriography, Thrombolytic Therapy, Em bolizations, Em bolectomy, IVC filter placement, Dilatations, Stent placement. Various: Arthrograms, Myelograms, Face

Reason for recall

The locking plate may be missing from the system, which may result in unexpected stand/table tilt movement that cannot be stopped by the user.

Distribution

US Nationwide domestic distribution, and worldwide international distribution.

Key facts

Status
Terminated
Initiation date
2019-02-28
Report date
2019-04-03
Termination date
2020-08-03
Voluntary/Mandated
Voluntary: Firm initiated
Location
Andover, MA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1055-2019