FDA recall Z-1074-2025

Getinge Usa Sales Inc · Class I · device

Product

Vaporizer Sevoflurane, Quik-Fil. Model Number: 6682285. Device is a component of Flow-i C20, Flow-i C30, Flow-i C40, Flow-c, Flow-e. For use during anesthesia.

Reason for recall

Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of respiratory tract, which in worst case may lead to lung edema and/or severe hypocalcemia, blistering, superficial ulceration, and/or hypomagnesemia.

Distribution

Domestic distribution to Kentucky and Connecticut.

Key facts

Status
Ongoing
Initiation date
2025-01-15
Report date
2025-02-19
Voluntary/Mandated
Voluntary: Firm initiated
Location
Wayne, NJ, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1074-2025