FDA recall Z-1078-2026

Elekta, Inc. · Class II · device

Product

MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)

Reason for recall

Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice

Distribution

US Nationwide distribution in the state of TX.

Key facts

Status
Ongoing
Initiation date
2025-12-10
Report date
2026-01-21
Voluntary/Mandated
Voluntary: Firm initiated