FDA recall Z-1096-2020

Microbiologics Inc · Class II · device

Product

Vaginal Verification Panel Ref 8208 Lot 8208-11

Reason for recall

Product 8208 Vaginal Verification Panel is a 6 pool kit and has contaminant Candida krusei in Pool 2 (subcomponent 6063). Candida krusei is an organism intended to be contained in Pool 6 of the kit. This should have no impact on patient test results as long as Gardnerella vaginalis was properly detected when using Pool 2 and/or users are strictly using this product for instrument verification/validation.

Distribution

Distributed in the US to 7 states: CA, KS, ME, NC, NY, OR, and UT. No Canada, Mexico or VA.

Key facts

Status
Terminated
Initiation date
2020-01-03
Report date
2020-02-12
Termination date
2020-09-16
Voluntary/Mandated
Voluntary: Firm initiated
Location
Saint Cloud, MN, United States

Primary source