FDA recall Z-1097-2025

Ethicon, LLC · Class II · device

Product

PERMA-HAND Silk Suture, REF: K872H, W723H, 623H, K833H; Perma-hand Silk Suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurosurgical procedures.

Reason for recall

Sutures may have an open seal on the primary packaging, due to a manufacturing issue on a specific machine, and if issue is undetected could lead to breach in sterility, could introduce pathogens to the patient and cause infection.

Distribution

Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, LA MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI and the countries of AT, AU, BE, CA, CH, CN, CZ, DE, DK, FI, FR, GB, HU, IE, JP, LU, NL, NO, PL, SE, SG, ZA.

Key facts

Status
Ongoing
Initiation date
2024-12-20
Report date
2025-02-12
Voluntary/Mandated
Voluntary: Firm initiated
Location
San Lorenzo, PR, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1097-2025