FDA recall Z-1099-2026

Thoratec LLC · Class II · device

Product

CentriMag Blood Pump, REF: 102953, 201-20003, 201-90010, 201-90016; CentriMag RVAD HDE KIT, REF: 104318; CentriMag Pre-connected Pack, REF: CMAEK00; CentriMag VAD KIT, REF: 201-51006, 201-51007, 201-90001 201-90001T; Blood Pump with CentriMag Acute Circulatory Support System for ECMO, REF: CMAEK01. CentriMag Blood Pump is a component to be used with CentriMag Acute Circulatory Support System

Reason for recall

Acute circulatory support system pump may not fit into the motor and reports received of the pump not being securely locked into the motor using the screw-in locking feature. Failure to properly lock the pump into motor my not cause pump not inserted alarm, and could lead to particulate, increased heat, anxiety, inconvenience, delay in hemodynamic/ECMO support, stroke, embolism, death.

Distribution

Worldwide - US Nationwide distribution including in the states of MN, PA, OK, FL, CA, OR, IL, NY, DE, GA, NJ, NC, WI, OH, AZ, AR, TX, TN, MO, MA, MT, KY, NE, WV, VA, ME, ND, AL, CO, DC, MS, IN, MI, CT, SC, WA, NM, MD, HI, ID, IA, LA, UT, RI, NV and the countries of Algeria, Australia, Austria, Bahrain, CANADA, Cayman Islands, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Hong Kong, Italy, Jordan, Kuwait, Lithuania, Macedonia, Malaysia, Mexico, Netherlands, Oman, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom.

Key facts

Status
Ongoing
Initiation date
2025-10-10
Report date
2026-01-21
Voluntary/Mandated
Voluntary: Firm initiated