FDA recall Z-1101-2025

Nobel Biocare · Class II · device

Product

. Nobel Biocare N1 TiUltra TCC RP 4.0x9mm . Nobel Biocare N1 TiUltra TCC RP 4.0x11mm . Nobel Biocare N1 TiUltra TCC RP 4.0x13mm Nobel Biocare N1 TiUltra TCC implants are intended for use as an endosseous dental implant in the maxilla or mandible for anchoring or supporting dental protheses to restore chewing function

Reason for recall

Due to manufacturing issue (unintended by-product of the injection molding process) a sharp pin that may puncture the blister packaging and breaching the sterile barrier of the finished product.

Distribution

U.S. Nationwide distribution in the states of CA, FL, IL, MA, NJ, NY, OH, PA, UT, VA and WA.

Key facts

Status
Ongoing
Initiation date
2025-01-10
Report date
2025-02-12
Voluntary/Mandated
Voluntary: Firm initiated
Location
Goteborg, Sweden

Primary source