FDA recall Z-1104-2020

Dexcom Inc · Class II · device

Product

Brand Name: Dexcom G6 CGM App for iOS Software Number: SW11677 Software Version: Version 1.4.0 up to corrected Version 1.4.4 GUDID Registered GTIN: 00386270000804

Reason for recall

It was reported that the user's low alarm feature on the iOS application were not properly alerting users when the user has enabled the Alert Schedule feature more than 30 days after installing the iOS application.

Distribution

US Nation-wide OUS: Austria; Belgium; Canada; Czechia; Denmark; Finland; Germany; Great Britain; Hong Kong; Ireland; Italy; Luxembourg; Netherlands; Norway; South Africa; Spain; Sweden, and Switzerland.

Key facts

Status
Terminated
Initiation date
2019-11-19
Report date
2020-02-19
Termination date
2023-08-22
Voluntary/Mandated
Voluntary: Firm initiated