FDA recall Z-1105-2023

Roche Molecular Systems, Inc. · Class II · device

Product

cobas SARS-CoV-2 & Influenza A/B Qualitative nucleic acid test for use on the cobas 5800/6800/8800 Systems; Material Number 09446125190

Reason for recall

The firm received customer complaints regarding false negative Influenza A (Flu A) results and late Flu A Target Ct values. A false negative influenza A result may lead to additional testing, psychological distress, and delays to targeted therapy for influenza as well as true diagnosis.

Distribution

Domestic distribution nationwide. Foreign distribution worldwide.

Key facts

Status
Ongoing
Initiation date
2022-12-21
Report date
2023-02-15
Voluntary/Mandated
Voluntary: Firm initiated