FDA recall Z-1106-2019

Randox Laboratories Ltd. · Class III · device

Product

Immunoassay Premium Plus Controls, Model Numbers IA3109 and IA3112 Product Usage: This product is intended for in vitro diagnostic use, as assayed quality control material to monitor the accuracy and reproducibility of analytes listed in the package insert

Reason for recall

The target value for uptake is incorrect on the value sheet for an analyzer. The target value for this analyzer is listed as 43.8%U and should be 39.4%U.

Distribution

The products were distributed to the following US states: WV and Puerto Rico.

Key facts

Status
Terminated
Initiation date
2019-02-07
Report date
2019-04-17
Termination date
2020-06-08
Voluntary/Mandated
Voluntary: Firm initiated
Location
Crumlin (North), N/A, Ireland

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1106-2019