FDA recall Z-1107-2021

Phadia US Inc · Class II · device

Product

EliA dsDNA Well - Product Usage: intended for the In-Vitro quantitative measurement of IgG antibodies directed to dsDNA in human serum and plasma as an aid in the clinical diagnosis of systemic lupus erythematosus (SLE). Material Number: 14-5500-01

Reason for recall

Decreased values for EliA ANA Positive Control when using the EliA dsDNA Well lot BFA37/0142 may potentially cause false positive results (increase in patient test results).

Distribution

US Nationwide distribution.

Key facts

Status
Terminated
Initiation date
2020-12-14
Report date
2021-02-17
Termination date
2022-08-04
Voluntary/Mandated
Voluntary: Firm initiated
Location
Portage, MI, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1107-2021