FDA recall Z-1108-2026

RAYSEARCH LABORATORIES AB · Class II · device

Product

Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 15.2.0, 15.1.3 Software Version: RayStation 2024A SP1, RayStation 2024A SP2 Product Description: Radiation Therapy Treatment Planning System Component: No

Reason for recall

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

Distribution

Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS

Key facts

Status
Ongoing
Initiation date
2025-11-28
Report date
2026-01-28
Voluntary/Mandated
Voluntary: Firm initiated
Location
Stockholm, Sweden

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1108-2026