FDA recall Z-1111-2026

ICU Medical, Inc. · Class II · device

Product

Tego Connector: REF: 011-D1000, 01C-D1000, 055-D1000, D1000; Conector Tego: REF: LAT-D1000

Reason for recall

Needle free access device intended as accessory to vascular access used in hemodialysis or accessory to intravascular administration sets has a silicone seal which may bulge at the top, separate from the body, tear; may result in fluid leaks, occluded fluid flow, inability to inject/withdraw blood product, delayed/interrupted therapy, exposure to biological contaminants, air infused into the body

Distribution

Worldwide - US Nationwide distribution including in the states of FL, AZ, RI, TN, KS, NJ, NY, VA, SC, GA, CA, MS, TX, OH, NC, MA, WA, IL, PA, NV, NH, LA, MI, NE, ME, MN, NM, OK, MD, IA, OR, SD, WI, MO, AL, WY, IN, KY, DC, UT, AR, WV, AK, ND, CO and the countries of BE, DE, CO, ES, AU, AR, SA, SI, IT, LK, MY, AT, FI, GB, TW, MX, NL, IE, CA, GR, ZA, SE, SK, NZ, LT, CH, HR, FR, JO, PE, SG, BR, PL, RO, HU, CZ, NO, AE, EE, CN, MA, HK.

Key facts

Status
Ongoing
Initiation date
2025-12-19
Report date
2026-01-28
Voluntary/Mandated
Voluntary: Firm initiated
Location
San Clemente, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1111-2026