# FDA recall Z-1112-2025

> **Beckman Coulter Inc.** · Class II · device recall initiated 2025-01-02.

## Product

Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay Systems

## Reason for recall

Identified reagent lots are experiencing a high rate of calibration failures with failure code of "Max Iterate" on Access 2 and UniCel Dxl Immunoassay Systems. Reporting of results may be delayed due to the time required to achieve a passing calibration curve.

## Distribution

Worldwide distribution - US Nationwide and the countries of Austria, Bahrain, Bangladesh, Belgium, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Croatia, Czechia, Denmark, France, Germany, Greece, Hungary, India, Ireland, Italy, Kazakhstan, Korea, Republic of, Malaysia, Mexico, Peru, Philippines, Poland, Puerto Rico, Romania, Slovakia, South Africa, Spain, Switzerland, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, Viet Nam.

## Key facts

- **Recall number:** Z-1112-2025
- **Recalling firm:** Beckman Coulter Inc.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2025-01-02
- **Report date:** 2025-02-12

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Brea, CA, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1112-2025