FDA recall Z-1113-2019

Siemens Medical Solutions USA, Inc · Class II · device

Product

syngo.via syngo.CT Cardiac Planning, Model Number 10496180 Product Usage: Syngo.CT Cardiac Planning and syngo CT. Cardiac Function are image analysis software packages for evaluating cardiac CT angiography (CTA) volume sets. The software packages are designed to support the physician in determining the functional parameters of the left and right ventricles, confirming the presence or absence of physicianidentified myocardial enhancement defects and evaluation documentation and follow-up of any such finding.

Reason for recall

There is a potential risk of a wrong measurement in the annulus plane during a TAVI planning procedure. This risk is due to a software issue found in the TAVI algorithm.

Distribution

US Nationwide Distribution in the states of: AK, AL, AR, AZ, CA, CT, FL, GA, IA, IL, KS, KY, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, and WI.

Key facts

Status
Terminated
Initiation date
2019-03-21
Report date
2019-04-17
Termination date
2022-01-11
Voluntary/Mandated
Voluntary: Firm initiated
Location
Malvern, PA, United States