FDA recall Z-1113-2026

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. · Class II · device

Product

Philips Azurion system; Software Version Number: R3.1;

Reason for recall

Philips has identified two (2) software issues affecting device systems that may result in loss of imaging (X-ray) functionality and/or loss of motorized movement, and/or incorrect image content and/or loss of data. Issue 1 - System remains in continuous restart mode after the start-up. Issue 2 - Longitudinal Position Error Applicable only to systems with Poly-G3 frontal stand.

Distribution

International distribution in the countries of AE, AR, AT, AZ, BE, BG, BH, BR, BS, CH, CL, CO, CR, CY, CZ, DE, DK, DZ, EC, EG, ES, GB, GR, GT, ID, IE, IL, IN, IQ, IT, JE, JP, LB, MA, MO, NI, NL, NO, PE, PL, RO, SA, SE, SV, TR, UA, VE, VN, ZA;

Key facts

Status
Ongoing
Initiation date
2025-12-12
Report date
2026-01-28
Voluntary/Mandated
Voluntary: Firm initiated
Location
Best, Netherlands

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1113-2026