FDA recall Z-1118-2026

ICU Medical, Inc. · Class II · device

Product

ICU Medical Plum Duo Infusion Pump, List Numbers: 400020401 (US), 400021001 (Philippines)

Reason for recall

Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.

Distribution

Worldwide distribution - US Nationwide and the country of Philippines.

Key facts

Status
Ongoing
Initiation date
2025-12-15
Report date
2026-01-28
Voluntary/Mandated
Voluntary: Firm initiated
Location
Lake Forest, IL, United States