FDA recall Z-1133-2019

BioFire Diagnostics, LLC · Class II · device

Product

FilmArray BCID Panel, IVD, Rx Only, BioFire Diagnostics, REF: RFIT-ASY-0126, and RFIT-ASY-0127, when used with the following blood culture bottles: Bottle Description/Part Number: BACT/ALERT FA Plus/410851; BACT/ALERT FN Plus/410852; BACT/ALERT PF Plus/410853

Reason for recall

Increased risk of false positive Enterobacteriaceae and Escherichia coli results on affected blood culture identification panels when using certain blood culture bottles.

Distribution

MA, NH, VT, CT, NJ, NY, PA, DE, DC, VA, MD, WV, NC, SC, GA, FL, AL, TN, MS, KY, OH, IN, MI, IA, WI, MN, SD, ND, MT, IL, MO, KS, NE, LA, AR, OK, TX, CO, WY, ID, UT, AZ, NM, NV, CA, HI, OR, WA, AK, BM

Key facts

Status
Terminated
Initiation date
2019-03-07
Report date
2019-04-24
Termination date
2022-08-24
Voluntary/Mandated
Voluntary: Firm initiated
Location
Salt Lake City, UT, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1133-2019