FDA recall Z-1135-2018

Smiths Medical ASD Inc. · Class II · device

Product

Medfusion Syringe Pump 3500 Series The Medfusion¿ Syringe Infusion Pumps are indicated for the following uses: 1) Administration of fluids requiring precisely controlled infusion rates including blood or blood products, lipids, drugs, antibiotics, enteral solutions and other therapeutic fluids; 2) By the following delivery routes: arterial, epidural, intravenous, intrathecal, subcutaneous, and enteral; 3)By the following delivery modes: continuous, volume/time, mass, body weight, intermittent, and bolus; 4) In critical care, anesthesia, neonatal, and pediatric applications or other healthcare settings where use of the syringe infusion pump can be monitored or supervised by a clinician; 5) Inside the MRI room mounted outside the 150 Gauss line and with shielded magnets of field strength of 1.5 Tesla.

Reason for recall

Certain Medfusion Syringe Pump Models, Series 3100, 3500, and 4000, may not recognize or may misidentify loaded medication syringes. The inability of a pump to recognize a syringe (i.e. the size of the syringe is unknown to the pump) results in an inability to complete pump programming. Misidentification of a syringe is where the pump misinterprets the syringe size.

Distribution

Worldwide distribution- US nationwide Distribution and to the countries of : Australia, Barbados, Canada, Hong Kong, Indonesia, Jamaica, New Zealand, Philippines, Saudi Arabia, Singapore, South Africa, Trinidad and Tobago, United Arab Emirates, and United Kingdom.

Key facts

Status
Terminated
Initiation date
2017-11-13
Report date
2018-03-28
Termination date
2025-09-16
Voluntary/Mandated
Voluntary: Firm initiated
Location
Minneapolis, MN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1135-2018