# FDA recall Z-1136-2021

> **Ortho-Clinical Diagnostics, Inc.** · Class II · device recall initiated 2021-01-14.

## Product

VITROS Chemistry Products VersaTips - Product Usage: used to dispense samples to MicroSlide assays and sample and reagent fluid for MicroWell assays. In addition, VersaTips are used to dispense R1 Reagent Fluid into cuvettes in combination with sample fluid for Micro Tips Assays.

## Reason for recall

A deformity affecting a specific lot of VITROS¿ Chemistry Products VersaTips. Deformed VersaTips results may be biased by up to +/-16% for both immunometric and competitive assays.

## Distribution

Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI and the country of Canada.

## Key facts

- **Recall number:** Z-1136-2021
- **Recalling firm:** Ortho-Clinical Diagnostics, Inc.
- **Classification:** Class II
- **Product type:** device
- **Status:** Terminated
- **Initiation date:** 2021-01-14
- **Report date:** 2021-03-03
- **Termination date:** 2022-12-27

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Rochester, NY, United States