# FDA recall Z-1136-2026

> **Sysmex America, Inc.** · Class II · device recall initiated 2025-12-15.

## Product

TS-10/TS-10H Tube Sorter, automatic sorting system for sorting barcoded sample tubes, Catalog Number BN168792, running software versions 00-17

## Reason for recall

Under specific conditions, the lift's electromagnetic holding force may disengage unexpectedly, allowing unintended downward movement of the lift.

## Distribution

US Nationwide.

## Key facts

- **Recall number:** Z-1136-2026
- **Recalling firm:** Sysmex America, Inc.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2025-12-15
- **Report date:** 2026-01-28

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Lincolnshire, IL, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1136-2026

## Citation

> AI Analytics. FDA recall Z-1136-2026. Retrieved 2026-07-27 from https://api.ai-analytics.org/recall/Z-1136-2026. Source: US FDA. Licensed CC0.

---

*[Dataset catalog](/datasets/) · [FDA recalls dataset](/datasets/fda-recalls) · CC0 1.0*