FDA recall Z-1141-2018

Becton Dickinson & Company · Class I · device

Product

BD Vacutainer¿ EDTA Pink Top Tube; Catalog Numbers: 1. 367842 (UDI: 50382903678421); 2. 367899 (UDI: 50382903678995); 3. 368049 (UDI: 50382903680493); 4. 368589 (UDI: 50382903685894)

Reason for recall

BD Vacutainer EDTA Lavender, Tan, and Pink Top Tubes and BD Vacutainer Lithium Heparin Green Top Tubes can cause an underestimation of lead in blood samples when used with Magellan Diagnostics LeadCare assays, employing the Anodic Stripping Voltammetry (ASV) methodology, or any other assay employing ASV methodology.

Distribution

Distributed in 46 states: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NY, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and Puerto Rico.

Key facts

Status
Terminated
Initiation date
2018-03-22
Report date
2018-03-28
Termination date
2022-07-25
Voluntary/Mandated
Voluntary: Firm initiated
Location
Franklin Lakes, NJ, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1141-2018