# FDA recall Z-1141-2019

> **Philips Medical Systems Nederlands** · Class II · device recall initiated 2019-02-19.

## Product

Foot Switches used with the following systems:   722001 	Allura Xper FD 10 C  722002	Allura Xper FD10  F  722003	Allura Xper FD10  722005	Allura Xper FD10/10  722006	Allura Xper FD20  722008	Allura Xper FD20 Biplane  722010	Allura Xper FD10  722011	Allura Xper FD10/10  722012	Allura Xper FD20  722013	Allura XPER FD20 BIPLANE  722014	Allura Xper FD10 OR Table  722015	Allura Xper FD20 OR Table  722019	Allura Xper FD10/10 OR Table  722020	Allura Xper FD20 Biplane OR Table  722022	Allura Xper FD10 OR Table  722023	Allura Xper FD20 OR Table  722024	Allura Xper FD10/10 OR Table  722025	Allura Xper FD20 Biplane OR Table  722026	Allura Xper FD10  722027	Allura Xper FD10/10  722028	Allura Xper FD20  722029	Allura Xper FD20/10  722031	Allura CV20  722033	Allura Xper FD10 OR Table  722034	Allura Xper FD10/10 OR Table  722035	Allura Xper FD20 OR Table  722036	Allura Xper FD20/10 OR Table  722038	Allura Xper FD20/20  722039	Allura Xper FD20/20 OR Table  722058	Allura Xper FD20/15  722059	Allura Xpe

## Reason for recall

Additional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or continuous inability of making live fluoroscopy images or exposures.

## Distribution

US Nationwide Distribution in the states of  CA, FL, MI, OR, PA, TX and WA.

## Key facts

- **Recall number:** Z-1141-2019
- **Recalling firm:** Philips Medical Systems Nederlands
- **Classification:** Class II
- **Product type:** device
- **Status:** Terminated
- **Initiation date:** 2019-02-19
- **Report date:** 2019-04-24
- **Termination date:** 2020-10-14

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Best, N/A, Netherlands

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1141-2019

## Citation

> AI Analytics. FDA recall Z-1141-2019. Retrieved 2026-06-04 from https://api.ai-analytics.org/recall/Z-1141-2019. Source: US FDA. Licensed CC0.

---

*[Dataset catalog](/datasets/) · [FDA recalls dataset](/datasets/fda-recalls) · CC0 1.0*