FDA recall Z-1143-2024

Voco GmbH · Class II · device

Product

Ionolux, Catalog (REF) Numbers: a) 2115, b)2117, c) 2118, d) 2119, e) 2120, f) 2122; dental cement

Reason for recall

Due to an error in the manufacturing process, it is possible that an incorrect plunger has been fitted to some capsules. This plunger does not contain the correct reactive component for Ionolux. When these capsules are activated, curing is significantly delayed and incomplete. Processing according to the instructions for use is not possible.

Distribution

Worldwide distribution: US distribution to states of: IA, NY, PA, WA; and OUS (foreign) to countries of: Belgium, Bulgaria, Denmark, Germany, France, Greece, Great Britain, India, Italy, Canada, Quatar, Lithuania, Netherlands, Northern Ireland, Norway, Poland, Portugal, Romania, Sweden, Switzerland, Slovakia, Spain, Czech Republic, Australia, Saudi Arabia, & Ireland.

Key facts

Status
Ongoing
Initiation date
2024-01-25
Report date
2024-02-28
Voluntary/Mandated
Voluntary: Firm initiated
Location
Cuxhaven, N/A, Germany

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1143-2024