FDA recall Z-1148-2020

Stryker Corporation · Class II · device

Product

Stryker, Omega 4.75 mm,. PEEK Knotless Anchor System, Single, REF 3910-500-471, UDI 07613327464818, Sterile R, RX Only CE 0197

Reason for recall

Complaints have been received regarding difficulty in removing the driver from the screw. High driver removal forces can lead to the screw pulling out still attached to the driver. This could result in an additional anchor insertion site in the bone being used.

Distribution

US: GA, NJ, NY, KY, MD, TX, UT, AZ, IL, ID, CO, OH, FL, AK, CA, NV, AR, ME, WI

Key facts

Status
Terminated
Initiation date
2019-07-19
Report date
2020-02-19
Termination date
2024-02-02
Voluntary/Mandated
Voluntary: Firm initiated
Location
San Jose, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1148-2020