FDA recall Z-1149-2021

Steris Corporation · Class II · device

Product

OT1000 Series Orthopedic Surgical Tables

Reason for recall

On January 7, 2021, STERIS identified that the glued bond between the beam and the leg spar assembly may begin to separate over time. This could result in unexpected movement of the affected leg spar if the separation occurred during a patient procedure.

Distribution

Domestic: AL, AK, AR, AZ, CA, CO, CT, DC, FL, GA, ID, IL IN, IW, KS, KY, LA, ME, MD, MI, MN, MO, NE, NV, NJ, NY, NC, ND, OH, OK, OR, PA, PR, SC, TX, UT, VT, VA, WA, WV, WI. Foreign: Canada, South Korea, Panama, Qatar, Thailand.

Key facts

Status
Terminated
Initiation date
2021-01-21
Report date
2021-03-03
Termination date
2021-11-10
Voluntary/Mandated
Voluntary: Firm initiated
Location
Mentor, OH, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1149-2021