FDA recall Z-1153-2024

Boston Scientific Corporation · Class II · device

Product

EndoVive Replacement Button Kit w/ ENFIT 28F 2.8 CM, Outer Box UPN M00509500 or M00509510.

Reason for recall

Affected batches of the kit contain Stoma Measuring Devices sized to measure for an 18/24F replacement button, but are labeled as being for a 28F. Use of defective product may lead to undersizing the button, which may cause embedding with erosion into the gastric wall, tissue necrosis, infection, sepsis, and associate sequelae.

Distribution

International distribution to the country of Japan only.

Key facts

Status
Ongoing
Initiation date
2023-12-19
Report date
2024-02-28
Voluntary/Mandated
Voluntary: Firm initiated
Location
Marlborough, MA, United States

Primary source