# FDA recall Z-1154-2020

> **Respironics California, LLC** · Class II · device recall initiated 2019-09-11.

## Product

Respironics RP-Touch Screen, V60/V680 Ventilator Serviced with Touchscreen RP Kit 453561511951 containing a touchscreen labeled as raw part number 1132318 Revision C

## Reason for recall

Ventilator touchscreen may become frozen and fail to respond to touch commands. If a patient requires a change in therapy, there is no warning associated with an unresponsive touchcreen, and therefore the clinician would have no prior indication of a touchscreen failure and would be unable to change ventilator settings. The ventilator will continue to function at the predetermined and accepted settings required to support the patient. A delay in therapy adjustments may result in hypercarbia and/or hypoxemia.

## Distribution

US Nationwide.    OUS (Foreign): Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Finland, France, Germany, Greece, Hong Kong, Iceland, India, Indonesia, Israel, Italy, Malaysia, Mexico, Netherlands, Portugal, Singapore, South Africa, Spain, Sweden, United Arab Emirates, United Kingdom, Viet Nam, Oman, Philippines, Saudi Arabia, Taiwan, Turkey, Russian Federation, Argentina, Ecuador, Poland, Guatemala, Peru, Viet Nam

## Key facts

- **Recall number:** Z-1154-2020
- **Recalling firm:** Respironics California, LLC
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2019-09-11
- **Report date:** 2020-02-19

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Carlsbad, CA, United States